In these recommendations you will find guidance on how to prime evercare® inline monoset with needle free connector.
1
Step 1
Unwrap the set. Close the clamp on side tubing, make sure the air vent is closed.
2
Step 2
Stick the side spike through the membrane into drug vial.
3
Step 3
Open the air vent on the side tubing.
4
Step 4
(A) Fill the drug vial by holding up the infusion bag and open the side clamp. Fill the drug vial half full, close the clamp or (B) attach a syringe to the needle free connector, fill the drug vial half full.
5
Step 6
To minimize a blocked filter: never infuse fluid into vial when inverted and never shake the vial when inverted.
6
Step 7
Close the air vent. Shake the vial side to side until the substance is dissolved.
7
Step 8
Open the air vent. Turn the drug vial upside down and raise it over the level of the infusion bag. Open the side clamp.
8
Step 9
When the drug is transferred to the infusion bag, close the clamp and the air vent. Leave the empty bottle on the side tubing.
Please note!
! See the package for complete IFU instructions for use, this is only a guide for users !
DO NOT OVER-TIGHTEN connections. DO NOT USE any instruments to tighten connections.