Priming evercare® inLine, transfusion set

In these recommendations you will find guidance on how to prime evercare® inLine Transfusion set.
1
Unwrap the IV-set and check wehether the roller clamp is open.
2
Change position of the roller clamp and control that the tubing is unaffected and close it.
3
Place the bag/container horizontal on a tabletop and insert ¾ of the length of the spike horizontally by using a turning motion.
4
Invert the container/- bag and hang it from the IV pole or similar.
5
Attach the tubing to the tube holder at the roller clamp.
6
Squeeze gently the drip chamber 3-5 times to fill half of the drip chamber (too aggressive squeeze can lead to mixed air fluid below the drip chamber).
7
Open the roller clamp until the complete tubing is primed.
8
Close the roller clamp.

Please note!

! See the package for complete IFU instructions for use, this is only a guide for users !

DO NOT OVER-TIGHTEN connections. DO NOT USE any instruments to tighten connections.

 