Priming evercare® inLine, unvented

In these recommendations you will find guidance on how to prime evercare® inLine infusion set when using semirigid container or infusion bag.
1
Priming evercare® inLine, unvented - Step 1

Step 1

Unwrap the IV-set and check that the roller clamp is open.
2
Priming evercare® inLine, unvented - Step 2

Step 2

Change position of the roller clamp and control that the tubing is unaffected and close it.
3

Step 3

Spike the semirigid container (A) or bag (B). Invert the container (A) or bag (B) and hang it from the IV pole or similar.
4
Priming evercare® inLine, unvented - Step 4

Step 4

Attach the tubing to the tube holder at the roller clamp.
5
Priming evercare® inLine, unvented - Step 5

Step 5

Squeeze gently the drip chamber 3-5 times to fill half of the drip chamber (too aggressive squeeze can lead to mixed air fluid below the drip chamber).
6
Priming evercare® inLine, unvented - Step 6

Step 6

Open the roller clamp untilthe complete tubing is primed.
7
Priming evercare® inLine, unvented - Step 7

Step 7

Closed the roller camp.

Additional information

Here you can find related products and PDF for download.


Products

REF: 10507-0210508-0110510-0110511-0110513-0210514-0110572-01105971059810599 

 

Download PDF

Please note!

See the package for complete IFU instructions for use, this is only a guide for users

DO NOT OVER-TIGHTEN connections. DO NOT USE any instruments to tighten connections

Tip: Glass and non fully collapsible containers: Keep the vent closed during spiking and priming. Only open the vent flap when venting is required.

How to prime evercare® inLine, unvented