Priming evercare® inLine, vented with semirigid container or bag

Priming recommendation to prime evercare® inLine infusion set when using semirigid container or infusion bag.
1
vented with semigrid container or bag - step 1

Step 1

Unwrap the IV-set and check whether the air vent cap is closed and the roller clamp is open.
2
vented with semigrid container or bag - step 2

Step 2

Change position of the roller clamp and control that the tubing is unaffected and close it.
3
InLine vented with bottle - step 3

Step 3

Spike the semirigid bag/container. Invert the bag/container and hang it from the IV pole or similar.
4
vented with semigrid container or bag - step 4

Step 4

Attach the tubing to the tube holder at the roller clamp.
5
vented with semigrid container or bag - step 5

Step 5

Squeeze gently the drip chamber 3-5 times to fill half of the drip chamber (too aggressive squeeze can lead to mixed air fluid below the drip chamber).
6
vented with semigrid container or bag - step 6

Step 6

Open the roller clamp until the complete tubing is primed.
7
vented with semigrid container or bag - step 7

Step 7

Closed the roller camp.